Optimized for Medical, Life Science, and Lab Automation

Posifa’s PLF2000 Series liquid flow sensors are a MEMS-based platform optimized for ultra-low liquid flow control in compact medical, life science, and lab automation equipment. With measurement ranges from 0–30 mL/min up to 0–200 mL/min, PLF2000 sensors help designers achieve stable, repeatable microflow delivery where even small deviations can affect assay integrity, dosing accuracy, or experimental outcomes.

FEATURES

  • Ultra-low flow ranges optimized at ≤ 200 mL/min
  • Fast response around 100 ms for tighter control in small-volume loops
  • Very low pressure drop with a straight, unobstructed flow path
  • No moving parts for long-term stability and fewer recalibration events
  • Push-to-connect fitting option to simplify manifold and tubing integration

 

BENEFITS

  • Improve dosing accuracy and reduce risk of overshoot
  • Reduce reagent waste and lower per-test cost
  • Achieve stable closed-loop control in small-volume systems
  • Minimize backpressure impact through low pressure drop
  • Extend service life with solid-state construction and no moving parts
  • Simplify pump, cartridge, and manifold integration with a compact, tubing-friendly form factor
  • Support both legacy and modern architectures with analog + I²C outputs

 

PLF2XXX APPLICATIONS

  • Lab automation reagent delivery that benefits from reduced waste and more consistent test results
  • Microfluidics and organ-on-chip perfusion where tight flow stability protects cell health
  • Flow cytometry and cell sorting fluidics where controlled delivery improves repeatability
  • Benchtop and production bioreactors where precise flow helps maintain target process conditions
  • Rodent infusion pumps for pharma research where accurate microdosing supports cleaner data
  • Infusion and micro-infusion pumps where preventing dosing overshoot is critical
  • HPLC and analytical instruments where stable low-flow measurements support consistent runs
  • Compact medical devices where predictable flow control improves safety margins
Model Typical flow range Outputs
PLF2003 0–30 mL/min Analog, I²C
PLF2040 0–40 mL/min Analog, I²C
PLF2005 0–50 mL/min Analog, I²C
PLF2010 0–100 mL/min Analog, I²C
PLF2020 0–200 mL/min Analog, I²C
Specifications Min Typ Max Unit Conditions
Ranges 30, 40, 50, 100, 200 ml/mm H20  
Analog Output 0.5   4.5 V  
Accuracy   ± 10   Reading 5 ~ 40 °C
Resolution   1 %   Reading  
Repeatability   2 %   Reading  
Response Time¹   100   ms  
Supply Voltage   5   VDC We recommend using 1% voltage regulator
Operating Current   22   mA  
Operating Temperature 5   40 °C  
Storage Temperature -25   80 °C  
Materials Nylon, silicone & food-grade waterproof coating (NSF/ANSI 61 compliant)